Herunterladen Inhalt Inhalt Diese Seite drucken

Installation Dimensions; Installing The Device - Endress+Hauser Nivotester FTR525 Anleitung

Vorschau ausblenden Andere Handbücher für Nivotester FTR525:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Nivotester FTR525
4.1.2
Orientation
The orientation can be in any position.
If the orientation is not vertical, the maximum operating temperature is reduced.
→ 4.1.1
4.1.3

Installation dimensions

Observe the width to be provided for the device of 45 mm (1.77 in).
45 (1.77)
19 Installation dimensions
4.2

Installing the device

Installation procedure:
S lide the upper top-hat rail clip upwards and the lower clip downwards until they reach
the points at which they click into place.
P osition the device on the top-hat rail from the front.
S lide the two top-hat rail clips together again until they click into place.
To remove the device, slide the top-hat rail clips upwards or downwards, as applicable (see 1.),
and remove the device from the top-hat rail. It is also sufficient to open one of the two top-hat
rail clips and tilt the device accordingly in order to remove it from the top-hat rail.
Endress+Hauser
118 (4.65)
Installation
mm (in)
49

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis