Herunterladen Inhalt Inhalt Diese Seite drucken

Connecting The Sensor; Connecting The Measuring Unit; Hart® Connection - Endress+Hauser iTEMP HART TMT162 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für iTEMP HART TMT162:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

TMT162
Endress+Hauser
4.2

Connecting the sensor

When connecting 2 sensors ensure that there is no galvanic connection between the
sensors (e.g. grounded duplex thermocouples). The resulting equalizing currents
distort the measurements considerably. In this situation, the sensors have to be
galvanically isolated from one another by connecting each sensor separately to a
field transmitter. The device provides sufficient galvanic isolation (> 2 kV AC)
between the input and output.
For sensor connection terminal layout please see Fig. 5. On two sensor inputs the following
connection combinations are possible:
Sensor 1:
RTD 2-wire
Sensor 2:
RTD 2-wire
Sensor 2:
RTD 3-wire
Sensor 2:
RTD 4-wire
Sensor 2:
TC connection
There are special cable entries available as accessories when connecting 2 sensors (not for
XP instrumentation). These can be found on  Page 85.
4.3

Connecting the measuring unit

NOTICE
Electronic parts may be damaged
Switch off power supply before installing or connecting the device. Failure to observe this
may result in destruction of parts of the electronics.
If the device has not been grounded as a result of the housing being installed, grounding
it via one of the ground screws is recommended.
®
4.3.1
HART
connection
®
If the HART
communication resistance is not built into the power supply, a 250 
communication resistor must be fitted into the 2-wire supply lines.
For connection hints, please take special notice of the documentation supplied by the
®
HART
Communication Foundation, specifically HCF LIT 20: "HART, a technical
overview".
Sensor 1:
RTD 3-wire
Yes
Yes
Yes
Yes
No
No
Yes
Yes
Sensor 1:
Sensor 1:
RTD 4-wire
TC connection
No
Yes
No
Yes
No
No
Yes
Yes
Wiring
65

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis