Commissioning
7.4.1.9

Parallel mode

Application
Sensor
2000
In order to prevent mutual interference from multiple microwave barriers operated in parallel,
up to five different channels can be set.
Options: "Channel 1" to "Channel 5"
• For details regarding the Soliwave microwave barriers, please see the associated
Technical Information TI01330F/97/EN and Operating Instructions
BA01804F/97/A2.
• I t is not necessary to carry out the channel setting again after changing the FQR57/
FDR57 devices, as the selected channel is automatically transmitted from the FTR525
to the FQR57 transmitter and the FDR57 transceiver.
7.4.2
Output adjustments
The following functions can be used for changing the settings for the signal outputs, including
the assignment and behavior of the relay/SSR and the current output.
7.4.2.1
Switching contact 1 assignment
Application
Output assignments
2000
This function is used to assign switching contact 1 (relay or SSR) to measuring channel 1
(limit, 1M), measuring channel 2 (motion, 2M) or the fault state.
Options: "Limit (1M)", "Motion (2M)", "Fault"
The option for the second measuring channel (motion, 2M) is only available for the corre-
sponding device versions.
7.4.2.2
Switching contact 2 assignment
Application
Output assignments
2000
This function is used to assign switching contact 2 (relay or SSR) to measuring channel 1
(limit, 1M), measuring channel 2 (motion, 2M) or the fault state.
Options: "Limit (1M)", "Motion (2M)", "Fault"
The option for the second measuring channel (motion, 2M) is only available for the corre-
sponding device versions.
64
Parallel mode
2100
2130
2200
2200
Switching contact 1
2210
Switching contact 2
2220
Nivotester FTR525
Endress+Hauser