Commissioning
The microwave barrier is optimally adjusted to match the application when the lower
detection limit (function 2112) corresponds to the minimum signal strength and the
upper detection limit (function 2113) corresponds to the maximum signal strength.
Application-specific
detection range
28 Optimal adjustment
Procedure for optimal adjustment:
M anually adjust the lower limit to match the application-specific minimum measured value.
Manually adjust the upper limit (function 2113) to match the application-specific maxi-
mum measured value.
The detection range can also be set to be lower or higher, for example if a turndown of the
signal output is desired.
7.4.1.3
Manual adjustment upper limit (UL) level limit detection
Application
Sensor
→
2000
This function can be used to change the upper limit of the set detections range using (for
increasing the set detection range 0 to 100 %) or (for reducing the set detection range 0 to
100 %). To assist with adjustment, the current measured value is shown alongside the
set detection range (→ 27 on page 61).
2 1 1 3
1M
4
3
29 Manual adjustment upper limit (UL)
1
Percentage value for the upper limit relative to the maximum possible detection range
2
Upper limit (adjustable using this function)
3
Current measured value
4
Set detection range (= 0-100% of signal output)
62
Limit detection
→
2100
2110
087
Manual adjustment UL
→
2113
1
2
Nivotester FTR525
Endress+Hauser