6. Commissioning, operation
Loads up to approx. 10 kΩ are harmless.
⇒
If the display exceeds the value set with dAC.1, then 1 V will be output.
⇒
If the display drops below the value set with dAC.0, then 0 V will be output.
⇒
In the event of an error (Err.1, Err.2, ----, etc.) an analogue signal slightly over 1 V will
EN
be output.
Jack connector wiring
WARNING!
Damage to property through incorrect jack connector
Using an incorrect jack connector or incorrect wiring can cause damage to
the hand-held.
▶
▶
▶
6.4.10 Alarm (AL.)
There are 3 possible settings: off = "AL.oFF", on = "AL.on", or on = "AL.no.So".
Under the following conditions, an alarm is given when the alarm function "AL.on" or "AL.
no.So" is active:
Value is below lower alarm limit "AL.Lo" or above upper alarm limit "AL.Hi".
■
Sensor error (Sens-Erro)
■
Low battery, 'bAt'
■
Err.7: System error
■
In the event of an alarm, the display flashes, with interface access the 'PRIO' flag is set.
6.4.11 Real-time clock (CLOC)
The real-time clock is needed for the time allocation of the logger data. If required, check
the settings.
Checking the real-time clock
1. Press the [SET/MENU] button for 2 seconds.
⇒
The main menu "SEt" is accessed.
2. Keep pressing the [SET/MENU] button until "SEt-CLOC" is displayed.
3. With the [TARA] button, select the "CLOC" parameters.
4. With the [MIN] or [MAX] buttons, enter the time.
5. With the [TARA] button, select the "dAtE" parameters.
6. With the [MIN] or [MAX] buttons, enter the day and month.
7. With the [TARA] button, select the "YEAr" parameters.
30
The 3rd connection must not be used.
Only stereo jack connectors are permitted.
Only use the original connection cable from WIKA.
WIKA operating instructions, model CPH62I0