6 Commissioning, Operation
6.9

Pressure connection to the sensors

The device is designed to be connected to the sensors model CPT6200 without a new calibration
being necessary. Therefore a great variety of replaceable sensors of 0 ... 100 mbar relative up to
0 ... 1000 bar pressure can be connected to the device.
Reference pressure sensors model CPT6200
For measurements of relative, vacuum or absolute pressure screw sensor to G1/2 pressure
connection or plug a plastic tube to a suitable adapter. Connection to CPH6300: Use sensor
connection cable.
Pressure measurements with two sensors (only 2-channel version)
Pressure differences of any sensor combinations can be measured by means of the calculation
sensor1 – sensor2 (DIF).
6.10 Alarm („AL.")
There are 3 possible settings:
off (AL.oFF), on with buzzer (AL.on), on without buzzer (AL.no.So).
Alarm is given in the following cases (if alarm active, AL.on or AL.no.So):
Lower alarm boundary (AL. Lo) under-run
Upper alarm boundary (AL. Hi) over-run
Sensor error
Low battery (bAt)
Err.7: system error (always with buzzer!)

6.11 Handling the tilt stand

Pull the tilt stand at the marking "open" to swing out the tilt stand
Pull the tilt stand at the marking "open" again to further swing out the tilt stand more.
Typical Applications:
Instrument fixed on a belt
WIKA Operating Instruction, Model CPH6300
Instrument placed on a table
instrument hanging on the wall
GB
27