WIKA CPH62I0 Betriebsanleitung Seite 24

Eigensicheres hand-held druckmessgerät
Vorschau ausblenden Andere Handbücher für CPH62I0:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6. Commissioning, operation
Menu
Parame-
ter
[MENU]
EN
Set Calibration: Adjustment of sensors
SEt
CAL
OFS.1
SCL.1
OFS.2
SCL.2
Set Alarm: Alarm function configuration
SEt
.
AL
AL.
[1,2,DIF]
AL.Lo
[1,2,DIF]
AL.Hi
[1,2,DIF]
Set Logger: Logger function configuration
SEt
Lo66
Func
CYCL
Lo.Po
Set Clock: Setting of real-time clock
SEt
CLOC
CLOC
dAtE
YEAr
1) If there is data in the logger memory, this menu item cannot be accessed. If these should be changed, the data must first be
deleted (see chapter 6.5 "Operation of the logger function").
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding
sensor on connection 2 then the settings are adopted.
3) If the automatic recording is activated, this menu item cannot be accessed. If these should be changed, the cyclic data logger
must first be stopped (see chapter 6.5 "Operation of the logger function").
4) If an alarm function limit is crossed (over or under), this is signalled by a "hooting" and a beeping.
24
Values
Meaning
or
The zero point of sensor 1 will be displaced by this value
Sensordep.,
e.g. -5.00 ...
to compensate deviations of the probe or the measuring
+5.00 mbar
instrument.
Zero point offset is deactivated (= 0.000)
oFF
The measuring scale of sensor 1 will be changed by this
e.g. -5.00
...+5.00 mbar
factor [%] to compensate deviations of the probe or the
measuring instrument.
Scale correction factor deactivated (= 0.000)
oFF
The zero point of sensor 2 will be displaced by this value to
Sensordep.,
e.g. -5.00 ...
compensate for deviations of the probe or the measuring
+5.00 mbar
instrument.
Zero point offset is deactivated (= 0.000)
oFF
The measuring scale of sensor 2 will be changed by this
e.g. -5.00
...+5.00 mbar
factor [%] to compensate deviations of the probe or the
measuring instrument.
Scale correction factor deactivated (= 0.000)
oFF
Alarm sensor is on; is indicated acoustically
on
Alarm sensor is on; is indicated acoustically
no.So
No alarm function
oFF
Min-Range ...
Min. alarm limit (not when AL. oFF , Sensor min. is the lower
display range limit of the connected sensor)
... AL.Hi
Max. alarm limit (not when AL. oFF , sensor max. is the upper
AL.Lo ...
... Max-Range
display range limit of the connected sensor)
Cyclic: Logger function "cyclic logger"
CYCL
Store: Logger function "individual value logger"
Stor
No logger function
oFF
Cycle time for cyclic logger [seconds]
1 ... 3600
Low-power logger with very low current supply (only with
on / oFF
cyclic logger and slow measurement)
HH:MM
Setting the time hours:minutes
TT.MM
Setting the time, day.month
YYYY
Setting the year
3)
4)
1)
WIKA operating instructions, model CPH62I0