8. Maintenance, cleaning and recalibration

8. Maintenance, cleaning and recalibration
Personnel: Skilled personnel
Protective equipment: Protective gloves and safety goggles
Tools: Spanner or torque spanner
For contact details see chapter 1 "General information" or the back page of
the operating instructions.

8.1 Maintenance

The model CPH62I0 hand-held pressure indicator is maintenance-free.
Repairs must only be carried out by the manufacturer.
This does not apply to the battery replacement.

8.2 Battery replacement

DANGER!
Danger to life from explosion
Through working in flammable atmospheres, there is a risk of explosion
which can cause death.
Only use the listed batteries, see chapter 3.6.3 "Permitted batteries"!
Do not use rechargeable batteries!
The instruments must not be opened in hazardous areas!
Only replace the batteries outside the hazardous area!
The battery cover must be closed and locked in place!
Ensure the correct polarity.
The cover of the battery compartment is located on the underside of the hand-held.
Procedure
1. Switch off the instrument and remove it from the leather case (Ex protective cover).
Slide the lid of the battery compartment, on the back of the instrument, downwards.
2. Remove the empty battery and pull off the connection cable.
3. Connect the connection cable to the new battery and then insert this back into the
battery compartment.
Make sure that the connection cable is connected with the correct polarity.
Only use approved batteries, see chapter 3.6.3 "Permitted
4. Slide the battery cover back on.
When closing the battery compartment make sure that the battery connection wires
are not jammed or damaged.
5. Insert the instrument back into the leather case (Ex protective cover).
WIKA operating instructions, model CPH62I0
batteries".
EN