10. Specifications
Reference pressure sensor model CPT62I0
Permissible ambient conditions
Medium temperature
EN
Operating temperature
Storage temperature
Relative humidity
Case
Material
Connection to the CPH62I0
Ingress protection
Dimensions
Weight
4) As an oxygen version, a flush diaphragm model is not available. In an oxygen version, the model CPT62I0 is only available in
overpressure ranges ≥ 0.25 bar (≥ 0.4 psi), with medium temperatures between -10 ... +50 °C (14 ... 122 °F) and using stainless
®
steel or Elgiloy

10.4 Safety-related characteristic values

Connection values CPH62I0-S1 or CPH62I0-S2
Max. voltage
Max. current
Max. power
Max. internal capacitance
Max. internal inductance
Power supply circuit CPT62I0
Max. input voltage
Max. input current
Max. input power
Max. internal capacitance
Max. internal inductance
46
4)
-20 ... +50 °C (-4 ... +122 °F)
-20 ... +50 °C (-4 ... +122 °F)
-40 ... +80 °C (-40 ... +176 °F)
0 ... 95 % r. h. (non-condensing)
Stainless steel
via 1.1 m (3.3 ft) sensor connection cable (plug-and-play);
Optional: Extension cable approx. 3.8 m (12.5 ft), overall cable
length approx. 5 m (16.4 ft)
IP65
IP67 when connected
See technical drawing
approx. 220 g (0.49 lbs)
wetted parts.
U
= DC 10.38 V
o
I
= 93 mA
o
P
= 240 mW
o
C
= 1,240 nF
o
L
negligible
o
U
= DC 10.4 V
i
I
= 100 mA
i
P
= 500 mW
i
= 600 nF
C
i
L
negligible
i
WIKA operating instructions, model CPH62I0

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cph62i0-s2Cph62i0-s1

Inhaltsverzeichnis