• Place the counter unit more than 0.5 m (20") away from a
high voltage source, large current source, large power
relay, etc.
• For installation of the counter unit, avoid a location
exposed to chips, cutting oil, or machine oil. If
unavoidable, take adequate countermeasures.
• Do not put a vinyl cover directly over the counter unit or
put it in a closed container.
• The ambient temperature should be in the range of 0 °C
to 40 °C (32 °F to 104 °F). Avoid exposure to direct
sunlight, hot air currents, or heated air.

1-3. Cautions on operation

Carry out the key operations or I/O (BCD, etc.) connections
and operations in line with the explanations given in the
appropriate sections. Failure to correctly operate this
device may result in a malfunction.
2
(E)
1-4. Instructions for connecting to
the measuring unit
• The LT10A/LT11A is a counter unit designed for the DT
series measuring unit.
• When the LT10A/LT11A is connected to a DT series and
turned on, the digits corresponding to the channel
connected may flash. This shows that initialization is
required to calibrate the new pairing of both device and
unit. Here you should move the measuring unit at least
1 mm, as shown in the figure, and then press the reset
key on the LT10A/LT11A corresponding to the channel.
The device will now return to its normal measuring state.
Note
Once this procedure has been carried out the digits will not
flash while the device is connected to that measuring unit,
even if the power is turned off.
R E
S E
T
1 mm or more
LT10A / LT11A Series
P