The Model 6506 Power-PRO XT cot is intended for use in the electromagnetic environment specified below . The
customer or the user of the Model 6506 Power-PRO XT cot should assure that it is used in such an environment .
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
flicker emissions
IEC 61000-3-3
Return To Table of Contents
1-108
EMC Information
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Group 1
Group 2
Cot: Class A
SMRT Charger
(6500-201-010): Class A
Cot: N/A
SMRT Charger
(6500-201-010): Class A
Cot: N/A
SMRT Charger
(6500-201-010): Complies
Electromagnetic environment - guidance
The Model 6506 Power-PRO XT cot uses RF energy only
for its internal function . Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment .
The Model
6506 Power-PRO XT cot must emit
electromagnetic energy in order to perform its intended
function . Nearby electronic equipment may be affected .
The Model 6506 Power-PRO XT cot is suitable for use in
all establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes .
The SMRT Charger is suitable for use in all establishments
other than domestic establishments other than domestic
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes .
Not applicable
Not applicable
6506-009-005 REV B