CAUTION (CONTINUED)
• A preventative maintenance program should be established for all Stryker EMS equipment . Preventative
maintenance may need to be performed more frequently based on the usage level of the product . Close attention
should be given to safety features including, but not limited to:
• Hydraulic power mechanism
• All electrical controls return to off or neutral position when released .
For additional maintenance information, see the preventative maintenance information .
• Improper maintenance can cause injury or damage to the product . Maintain the cot as described in this manual .
Use only Stryker approved parts and maintenance procedures . Using unapproved parts and procedures could
cause unpredictable operation and/or injury and will void the product warranty (see
• Failure to use authorized parts, lubricants, etc . could cause damage to the cot and will void the warranty of the
product .
• Hydraulic lines, hoses, and connections can fail or loosen due to physical damage, kinks, age, and environment
exposure . Check hoses and lines regularly to avoid damage to the cot . Check and tighten loose connections .
• Do not tip the cot onto its load wheels and actuate the product as this will allow air to enter the hydraulic system .
• Do not lubricate the bearings in the X-frame as it will degrade the performance of the cot and may void its warranty
(see
page
1-106) .
• The cot retaining post comes preconfigured for an X-frame cot, if the fastener has been configured for an H-frame
style cot, the cot retaining post must be adjusted to accommodate the fastener .
www.stryker.com
Summary of Safety Precautions
6506-009-005 REV B
page
1-106) .
Return To Table of Contents
1-15