Table 4
Recommended protective distances between portable or mobile HF telecommunication devices
and the Suprasorb
®
The Suprasorb
CNP P3 therapy unit is intended for operation in an electromagnetic
®
environment as described below in which HF interference is controlled.
The customer or user of the Suprasorb
interference by ensuring the minimum distance is maintained between portable and mobile
HF telecommunication devices (transmitters) and the Suprasorb
depending on the output power of the communication device, as specified below.
Nominal capacity
of transmitter
0.01
0.1
1
10
100
For transmitters with maximum nominal power not given in the table above, the recommended protective distance d in metres (m)
can be calculated using the formula in the respective column, where P is the transmitter's maximum nominal power in watts (W) in
accordance with the transmitter manufacturer's specifications.
NOTE 1: The higher frequency range applies at 80 MHz and 800 MHz.
NOTE 2: These guidelines may not be applicable in all cases. The transmission of electromagnetic parameters is influenced by the
extent to which they are absorbed and reflected by buildings, objects and people.
CNP P3 therapy unit.
®
Protective distance depending on transmitter frequency in m
150 kHz to 80 MHz
3.5
3
5 ,
d
[
]
P
=
V
1
0.12
0.37
1.17
3.69
11.67
CNP P3 therapy unit can help avoid electromagnetic
80 MHz to 800 MHz
3.5
3
5 ,
d
[
]
P
=
E
1
0.12
0.37
1.17
3.69
11.67
CNP P3 therapy unit –
®
800 MHz to 2.5 GHz
7
d
[
]
P
=
E
1
0.23
0.74
2.33
7.38
23.33