7 Cleaning and disinfection
• Clean the device regularly and disinfect the surfaces of the device and its accessories at
least once a week.
• The device and accessories can be wiped with a damp, lint-free cloth.
When the device has been in use for an extended period, the plastic components of the housing
may exhibit slight discolouration. This does not have any effect on the functioning of the device.
If liquids come into contact with the inside of the device, the unit must be checked by L&R or a
service partner authorised by L&R.
Avoiding infections
Infectious and pathogenic microbes in the exudate can pose health risks.
• Wear appropriate disposable gloves when replacing the exudate pouch.
• Replace the exudate pouch including the exudate pouch tube in accordance
with the respectively applicable hygiene regulations, at least every week,
before every new patient or when full.
• Cleaning is required whenever there is a new patient. Cleaning means that all
CAUTION
the single-use products are disposed of and the Suprasorb
and its accessories are disinfected by wiping. The housing of the
Suprasorb
authorised by L&R.
Cleaning the device
• Switch off the device prior to cleaning/disinfection by disconnecting the
power supply from the mains power and then from the power socket on the
Suprasorb
CAUTION
• Disconnection from the power supply is only guaranteed when the charging
cable is removed from the charging port.
Disinfecting the device
• We recommend wearing appropriate disposable gloves when performing
disinfection.
• Aggressive cleaning agents may damage the Suprasorb
and its accessories. Do not use any products containing acetone, or disinfectants
with a pH level greater than 9 or ethanol content greater than 35%.
• Follow the manufacturer's instructions for use for the disinfectants used, especially
PLEASE NOTE
with regard to the compatibility with the materials and surfaces as well as the
concentration specifications.
• Disinfection during the therapy is not expressly required when using the device
on a single patient at home.
• Disinfection is generally required when the device is used in an inpatient setting.
Use of the carry bag
The Suprasorb
is intended for mobile use by a patient. It should not be used for other patients, but be
CAUTION
disposed of after use.
CNP P3 therapy unit must only be opened by L&R or a service partner
®
CNP P3.
®