4.8 Confirming the therapy setting

Once the setting reflects the desired therapy,
confirm with
. You are automatically
returned to the therapy screen, where you
can now see the selected therapy settings.
If you do not wish to make any changes to
the therapy settings, you can return with

4.9 Starting the therapy and seal level

Regular patient monitoring
• Careful and regular patient monitoring is absolutely essential, particularly in the case of
heavily exuding wounds, wounds with viscous or heavily sedimenting exudate and
where there is a risk of severe bleeding (see Section 1.4.4 "Precautionary measures").
• In this situation, special attention should be given to exudate accumulated in the
WARNING
wound dressing and the correct positioning of the wound dressing.
• The monitoring interval must be determined based on the wound situation,
the patient's general condition and the sensitivity settings.
If bleeding occurs
The Suprasorb
the wound drainage system and to suction exudate.
• The Suprasorb
pooling of blood in the wound.
• For this reason, you should not rely on the Suprasorb
system alone. The therapy must be regularly monitored and the exudate regularly
WARNING
checked for red discolouration so as to identify any risk of severe deterioration in the
patient's condition.
• Should you observe bright red blood in the tubes or exudate pouch, the therapy
must be discontinued immediately. Leave the wound dressing in place for now,
stem the bleeding and seek medical assistance.
If you press
The screen goes green and the current pressure is shown.
The seal level can be seen on the right at the bottom of the screen.
It serves as a guide for how tightly the dressing is sealed.
.
CNP P3 therapy unit was designed to generate negative pressure in
®
CNP P3 therapy unit was not designed to detect bleeding or
®
now, you will start the therapy.
In this example, the
"Continuous" therapy mode
is selected, the negative
pressure is set to -80 mmHg
and the sensitivity is set to
"Medium".
CNP P3 therapy unit's alarm
®