Location
Multiplication factor 1.0
Table 1
Guidelines and manufacturer's declaration – Electromagnetic emissions
The Suprasorb
CNP P3 therapy unit is intended for operation in an electromagnetic
®
environment as described below.
The customer or user of the Suprasorb
such an environment.
Measurement of
emitted interference
HF emissions in
accordance with CISPR 11
HF emissions in
accordance with CISPR 11
Harmonic current
emissions in accordance
with IEC 61000-3-2
Emission of voltage
fluctuations/flickers
in accordance with
IEC 61000-3-3
0 metres
500 metres
above sea level
above sea level
0.94
CNP P3 therapy unit must ensure that it is operated in
®
Conformity
Group 1
Class B
Class A
Conforms
1,000 metres
above sea level
0.88
Electromagnetic environment – guidance
The Suprasorb
CNP P3 therapy unit only uses
®
HF energy for its internal workings.
As such, its HF emissions are very low and it is
unlikely that adjacent electronic devices will be
affected.
The Suprasorb
CNP P3 therapy unit is
®
intended for use in all facilities including
residential buildings and those directly connected
to the mains power grid which also supplies
residential buildings.
2,000 metres
above sea level
0.78
155