Table 3
Guidelines and manufacturer's declaration – Electromagnetic immunity
The Suprasorb
CNP P3 therapy unit is intended for operation in an electromagnetic
®
environment as described below.
The customer or user of the Suprasorb
that it is operated in such an environment.
Immunity
tests
Conducted HF
disturbances in
accordance with
IEC 61000-4-6
Radiated HF
disturbances in
accordance with
IEC 61000-4-3
NOTE 1: The higher frequency range applies at 80 MHz and 800 MHz.
NOTE 2: These guidelines may not be applicable in all cases. The transmission of electromagnetic parameters is influenced by
the extent to which they are absorbed and reflected by buildings, objects and people.
a In theory, it is not possible to determine the field strength of stationary transmitters, such as base stations of radio
phones and land mobile radio services, CB radio stations, AM and FM radio and TV stations in advance with accuracy.
A site study of the prevalent electromagnetic phenomena should be taken into consideration for the determination of the
electromagnetic environment with regard to stationary transmitters. If the field strength measured at the location at which the
Suprasorb
CNP P3 therapy unit is used exceeds the above conformity levels, the device should be monitored to verify that it
®
functions as intended. If unusual performance characteristics are observed, additional measures such as a modified alignment
or a different location for the device may prove necessary.
b The field strength should be less than 3 V/m over the frequency range from 150 kHz to 80 MHz.
®
Test
Conformity
level
level
3 V
3 V
3 V/m
80 MHz to
3 V/m
2.5 GHz
CNP P3 therapy unit must ensure
Electromagnetic environment –
guidance
Portable and mobile wireless devices should
not be used closer to the Suprasorb
therapy unit, including all cables, than the
recommended protective distance calculated
using the applicable equation for the
transmission frequency.
Recommended protective distance:
3.5
3
5 ,
d
[
]
=
V
1
3.5
3
5 ,
d
[
]
=
E
1
7
d
[
]
=
E
1
where P is the maximum nominal capacity of
the transmitter in watts (W) in accordance with
the transmitter manufacturer's specifications
and d is the recommended protective distance
in metres (m). The field strength of stationary
transmitters should be less than the conformity
level at all frequencies according to an on-site
(a) study. (b) Interference is possible in the
vicinity of devices bearing the following symbol:
P
P
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz
CNP P3
®
157