Development of seal level:
Start phase:
Negative pressure builds up. The LED bar at the top of the device is not yet activated.
Therapy phase:
Following the start phase, one of the following screens shows that the dressing is sealed and the
therapy functioning. The LED bar is green.
It is not possible to change the therapy settings whilst the therapy is running.
You need to interrupt the therapy by pressing
PLEASE NOTE
Interruption of therapy
Should it prove necessary to interrupt the therapy and switch the Suprasorb
therapy unit to Pause mode, this interruption should not be for longer than two hours
at a time within a period of 24 hours. The therapy must be continued after max. two
WARNING
hours. If the therapy is interrupted for a longer period, the dressing must be changed
before the therapy can be started again.
Unlocking the screen
Please note that the screen lock may activate during operation.
If the screen goes dark, you need only touch it to make it visible again. If you want to make
changes, press
until
appears to unlock the screen.
to change the settings.