Reference Information
Test frequency
(MHz)
2450
5240
5500
5785
NOTICE
If necessary to achieve the immunity test levels, the distance between the transmitting an-
tenna and the ME device or ME system can be reduced to 1 m. A testing distance of 1 m is al-
lowed by IEC 61000-4-3.
A.5 Training
After having been instructed and having read the operating instructions, training on the device is not es-
sential.
If there is a need, however, further training can be provided. Please contact PHYSIOMED ELEKTROMEDIZIN
AG.
A.6 Settings
In the Setting menu, you can set the brightness of the screen and switch the dialog language.
You can access the basic settings menu from the welcome screen which is displayed as soon as you insert
the therapy card.
®
DEEP OSCILLATION
Personal
Frequency band
Wireless service Modulation
(MHz)
DECT
LTE Band 1, 3, 4,
25
UMTS
2400...2570
Bluetooth
WLAN
802.11 b/g/n
RFID 2450
LTE Band 7
5100...5800
WLAN 802.11 a/
n
Max. power
(W)
Pulse modulation
2
217 Hz
Pulse modulation
0.2
217 Hz
PHYSIOMED ELEKTROMEDIZIN AG
Distance (m)
Immunity test
level (V/m)
0.3
28
0.3
9
27