8 Individual Programs (only DEEP OSCILLATION Personal Pro)
Chapter 8 Individual Programs (only DEEP OSCILLATION
Personal Pro)
Individual programs allow you to adjust therapy parameters such as duty cycle and frequency. The corres-
ponding INDIVID menu is only available for DEEP OSCILLATION
the DEEP OSCILLATION
You have two options for individual treatment:
INDIVID Sequence
INDIVID Manual
How to perform an INDIVID SEQUENCE treatment
(1)
Choose INDIVID in the side menu to get to the individual programs:
Individual Program - Page 1 - Selection of Parameters
(2)
Click again on INDIVID to get to the first page.
(3)
Click again on INDIVID. You can now rotate the Data Selector <2> to select the treatment dura-
tion, the duty cycle (e.g. 3:1), the frequency (e,g. 200 Hz) and the treatment duration of the three
phases. Move the cursor, which is indicated by a horizontal bar, to the desired parameter.
To select the value of a parameter, first press the Data Selector <2> to activate the parameter. Then
rotate the Data Selector <2> to select the desired value.
(4)
When you increase the intensity, the currently active phase is indicated by a black dot. The sequence
is completed phase after phase. Once the intensity is up you can only change the duty cycle and the
frequency. The duration within a sequence cannot be changed. Still, you can abbreviate (but not ex-
tend) the duration of the ongoing treatment.
(5)
Click on the symbol < in the top row of the Display <1> to return to the overview page.
®
DEEP OSCILLATION
Personal
®
Personal Basic version.
PHYSIOMED ELEKTROMEDIZIN AG
®
Personal Pro, and cannot be selected in