Reference Information
DEEP OSCILLATION
domestic health care.
A.4.1 Environment Emission Limits
The device is intended for operation in an environment as specified below. The customer or the user of the
device should ensure that it is operated in such an environment.
Phenomenon
Power-induced and radiated interference
emissions
Distortions caused by harmonics
Voltage fluctuations and flicker
A.4.2 Immunity
The device is intended for operation in an environment as specified below. The customer or the user of the
device should ensure that it is operated in such an environment.
Phenomenon
Discharge of static electricity
and connection to patients
High-frequency electromagnetic
fields
High-frequency electromagnetic
fields in the immediate vicinity
of wireless communication
devices
Magnetic fields with energy-
technical nominal frequencies
A.4.3 AC connection for the supply input
The device is intended for operation in an environment as specified below. The customer or the user of the
device should ensure that it is operated in such an environment.
Phenomenon
Fast transient electrical disturb-
ances/bursts
Surge voltages wire vs. wire
Surge voltages wire vs. earth
Wire-conducted disturbances,
®
DEEP OSCILLATION
Personal
®
Personal had been tested for usage in professional health care facilities and areas of
Professional health care facility
Class B, Group 1 (according to CISPR 11)
Refer to IEC 61000-3-2
Refer to IEC 61000-3-3
EMC basic standard or test
method
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-3
EMC basic standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
PHYSIOMED ELEKTROMEDIZIN AG
Area of domestic health care
Class B, Group 1 (according to CISPR 11)
Refer to IEC 61000-3-2
Refer to IEC 61000-3-3
Immunity test level for pro-
fessional health care facility
± 8 kV contact discharge
± 2 kV, ± 4 kV, ± 8 kV , ± 15
kV air discharge
3 V/m
80 MHz to 2,7 GHz
80 % AM at 1 kHz
Test specifications
Refer to
for immunity and covers
to HF wireless communic-
ation equipment
on page 26
30 A/m 50 Hz or 60 Hz
Immunity test level for pro-
fessional health care facility
± 2 kV 100 kHz repetition rate
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2 kV
3 V 0.15 MHz - 80 MHz 6 V in
Immunity test level for areas
of domestic health care
± 8 kV contact discharge
± 2 kV, ± 4 kV, ± 8 kV , ± 15
kV air discharge
10 V/m
80 MHz to 2,7 GHz
80 % AM at 1 kHz
Test specifications
Refer to
for immunity and covers
to HF wireless communic-
ation equipment
on page 26
30 A/m 50 Hz or 60 Hz
Immunity test level for areas
of domestic health care
± 2 kV 100 kHz repetition rate
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2 kV
3 V 0.15 MHz - 80 MHz 6 V in