14.12 Metrological Data
14.12.1 Models BCA324 | BCA224 | BCA124 | BCA64
Model
Scale interval (d)
Maximum capacity (Max)
Repeatability
At 5%, typical value
At approx. the maximum load, typical value
Linearity deviation
Limits
Typical value
Sensitivity drift between +10°C and +30°C
Tare maximum capacity (subtractive)
isoCAL (only for I-1x models):
Temperature change
Time interval
Only for models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Minimum initial weight according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weight
Typical minimum initial weight
Typical measurement time
Typical stabilization time
Recommended calibration weight
External calibrated test weight
Accuracy class in accordance with
OIML R111-1
Unit
mg
0.1
g
320
mg
0.08
mg
0.1
± mg
0.3
± mg
0.06
± ppm/K
1.0
<100% of maximum capacity
K
1.5
h
4
I
BC-AA
mg
1
mg
10
g
0.082
g
0.16
s
≤2.0
s
≤1.5
g
200
E2
Entris
II Advanced Line Operating Instructions
®
Technical Data
0.1
0.1
220
120
0.08
0.08
0.1
0.1
0.2
0.2
0.06
0.06
1.0
1.0
1.5
1.5
4
4
I
I
BC-AB
BC-AB
1
1
10
10
0.082
0.082
0.16
0.16
≤2.0
≤2.0
≤1.5
≤1.5
200
100
E2
E2
0.1
60
0.08
0.1
0.2
0.06
1.0
1.5
4
I
BC-AB
1
10
0.082
0.16
≤2.0
≤1.5
50
E2
53