8.8
Printing Weight Result With ID
Identifiers, e.g., device ID, batch ID, and sample ID, can
be assigned to the device, the sample, and to a batch.
The ID numbers are exported during ISO / GLP printout.
Requirements
− The ISO / GLP printout is activated in the "Settings" /
"Printout" / "Manual print" / "ISO / GLP printout"
menu.
− The device ID is set in the "Settings" / "Printout" /
"IDs" / "Set device ID" menu.
− A printer or a PC is connected and configured.
Procedure
Carry out weighing.
t
Press the [Start GLP] button.
t
If the "Batch ID" function is activated: Enter a batch
t
ID.
If the "Sample ID" function is activated: Enter a
t
sample ID.
The GLP header is printed with the device ID, the
y
batch ID, the sample ID, and the current weight
value.
To exit the GLP printout: Press the [Exit GLP] button.
t
The GLP footer is printed.
y
Cleaning and Maintenance
9
Cleaning and
Maintenance
9.1
Detaching the Analytical Draft
Shield
Material:
1 soft support base
Requirements
The device is disconnected from the power supply.
Procedure
Place the side panels and the upper panel on a soft
t
surface.
If an analytical balance is to be cleaned: Remove the
t
weighing pan and the shield disk.
If a precision balance is to be cleaned: Remove the
t
weighing pan.
Entris
II Advanced Line Operating Instructions
®
To detach the handles
t
on the side panels
and the upper panel:
Loosen the screws on
the inner sides of the
handles by turning.
CAUTION Glass
t
breakage due to
incorrect handling!
Slide the side panels
t
and upper panel all
the way back and out
of the guide rails.
43

Werbung

loading