Cleaning and Maintenance
9.2
Cleaning the Device
Procedure
Disconnect the device from the power supply.
t
Only use suitable cleaning agents and cleaning
t
procedures and observe the product information for
the cleaning agent used (for cleaning agent, see
Chapter "14.7 Cleaning Agent", page 51).
Remove dust and powdery sample residue with a
t
brush or hand-held vacuum cleaner.
Wipe the device and the associated components
t
with a slightly damp cloth. Use a mild soapy solution
or a suitable cleaning agent for more severe
contamination.
9.3
Installing the Analytical Draft
Shield
Procedure
If an analytical balance is used: Put the shield disk
t
and the weighing pan into place.
If a precision balance is used: Position the weighing
t
pan.
Connect the device to the power supply (see
t
Chapter 6.2, page 37).
44
Entris
II Advanced Line Operating Instructions
®
Slowly slide the side
t
panels and the upper
panel into the frame rails.
Hold the handles on the
t
side panels or the upper
panel from the outside
and fasten to the inner
side with the screws.
9.4
Maintenance Schedule
Interval
Component
Regularly;
Device
depending on
the operating
conditions
9.5
Software Update
For a software update, contact Sartorius Service.
Action
Perform a functional
test of the device.
Contact Sartorius
Service for this.