Operating Concept
4.11.2
Parameters in the "Settings / Date and Time" Menu
Parameters
Settings
Date
Date format
DD-MMM-YY
MMM-DD-YY
YY-MM-DD
(ISO)*
Time
Time format**
24h*
12h (AM/PM)
* Factory setting
** Only active if "YY-MM-DD (ISO)" is not selected under "Date format".
4.11.3
Parameters in the "Settings / Cal./Adj. Settings" Menu
Parameters
Settings
isoCAL
Off
Info, manual
start
Automatic*
Calibration
report
* Factory setting
22
Entris
II Advanced Line Operating Instructions
®
Explanation
Defines the date.
Sets the date display format to DD.MMM.YY
Sets the date display format to MMM.DD.YY
− Sets the date display format to ISO FORMAT YYYY-MM-DD.
− Sets the time to 24-hour mode.
Sets the time.
Sets the time to 24-hour mode.
Sets the time to 12-hour mode.
Explanation
Switches the isoCAL function off.
If the balance needs to be calibrated and adjusted: The [isoCAL] button is
highlighted orange. The isoCAL function must be manually triggered with the
[isoCAL] button.
Activates the automatic calibration and adjustment function as soon as
predefined time intervals or temperature values are exceeded.
Displays max. 99 calibration reports per calendar day.
Displays max. 30 calendar days.