81
Anhang
Guidance and manufacturer's declaration – electromagnetic immunity
®
UNIDRIVE
S III model 20 7010 20-1 is intended for use in the electromagnetic environment specified below.
The customer or user of the UNIDRIVE
Immunity test
EN/IEC 60601 test level
Electrostatic discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
Electrical fast transient/burst
± 2 kV for power supply lines
IEC 61000-4-4
± 1 kV for input/output lines
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to earth
Voltage dips, short
interruptions and voltage
<5 % U
variations on power supply
(>95 % dip in U
input lines IEC 61000-4-11
for 0.5 cycle
40 % U
(60 % dip in U
for 5 cycle
70 % U
(30 % dip in U
for 25 cycle
<5 % U
(>95 % dip in U
for 5 s
Power frequency (50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Appendix
Table 2
®
S III model 20 7010 20-1 should ensure that it is used in such an environment.
Compliance level
± 6 kV contact
± 8 kV air
2 kV/1 kV
1 kV/2 kV
0 V, 10 msec
*
T
)
T
92 V, 100 msec
T
)
T
168 V, 500 msec
T
)
T
0 V, 5 sec
T
)
T
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a typical commerci-
al or hospital environment.
Mains power quality should be that of a typical commerci-
al or hospital environment.
Mains power quality should be that of a typical commerci-
al or hospital environment.
If the user of the equipment or system requires continued
operation during power mains interruptions, it is
recommended that the equipment or system be powered
from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels cha-
racteristic of a typical location in a typical commercial or
hospital environment.