5.3 Digital calibration of analogue inputs
Step
To do
1
Select operation level "Service" from main menu, then select
address 910 "Channel calibration activate".
- Acknowledge selection with
- Enter service code "7049"
- Acknowledge selection of address 910 again using
2
Select channel
- Select channel with
- Channel 1: 920 ... 925
Channel 2: 930 ... 935
...........
3
Calibrate channel
- Follow the display commands (eg. connect 0 V/ push
- Using
4
End operation
- Operate
go to all calibration points and store values !