5 Problems and solutions
5.2 Adjust line recorder plot module
Step
To do
1
Select operation level "Service" from main menu, then select
address 910 "Channel calibration activate".
- Acknowledge selection with
- Enter service code "7049"
- Acknowledge selection of address 910 again using
(Display shows: Channel cal. deactivate).
Select channel
- Select channel with
Channel 1: Address 926,
Channel 2: Address 936,
Channel 3: Address 946
2
- Channel x: 70% adjust
- Acknowledge with
- Adjust channel to 70 %. Push button
pen left, push button
- Acknowledge adjustment end with
- Operate