4th. menu level (only change anything, if you know what you are doing!)
To access the 4th. menu level got to the 3rd. menu level and choose func�on CoDE. Enter the
access code (default: 3) with
Press key to scroll through the menu func�ons.
F
Again press keys and
F
menu.
TSET Set point of the thermostat switch
This value is set by the manufacturer and should not be changed.
(Adjust with
HY
Hysteresis of the thermostat switch
This value is set by the manufacturer and should not be changed.
(Adjust with
STBY Ini�al wai�ng procedure
OFF value: When it is started up, the calibrator immediately run to the last set point value
chosen before turning off.
ON value: When it is started up, the calibrator goes on the wai�ng posi�on and „STBY"
flashes on the display. Press any key in order to move it from the wai�ng posi�on and to
choose the desired Set Point value.
COD1 Indica�on of the access code for the 3rd. menu level
The factory set value is 2. The value can only be modified through serial communica�on,
write the desired value between 0 and 99 in variable 13 (see chapter 12).
COD2 Indica�on of the access code for the 4th. menu level
The factory set value is 3. The value can only be modified through serial communica�on,
write the desired value between 0 and 99 in variable 20 (see chapter 12).
DRUCK
&
TEMPERATUR
Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
LR-
Cal
LR-
Cal
keys, then press key and
▲▼
at the same �me or wait for 20 seconds for ge�ng back to the primary
▲
keys and confirm with key)
▲▼
keys and confirm with key)
▲▼
DT
PYROS 140-1L +
LR-
Cal
PYROS 375-DNV +
LR-
F
▲
E
E
-Info@Leitenberger.de •
www.druck-temperatur.de
63
PYROS 140-2L
MANUAL
Cal
PYROS 650-DNV
key at the same �me.