• Insert the sensor of the thermostat into a suitable
hole of an inserted sleeve.
• Connect the thermostat to the two sockets on the
front of the device using the cables supplied.
• Switch on the calibrator.
• Enter with the keys
to the T.
min
value. The value must be at least 10% lower than
the thermostat intertven�on point.
• Press the key
E
(Enter).
• Press the key (Func�on) to select func�on
F
Use the keys
and
▼
the thermostat. The value must be higher than the
interven�on point.
• Press the key (Enter).
E
• The thermostat switch temperature is expected to be
found within T.
min
• Press key (Func�on) to select the func�on
F
and
to enter the hea�ng slope rate value.
▼
▲
Low values are preferable for a more accurate test (e.g. value less than 1°C per minute is
preferable).
• Press the key (Enter).
E
The Switch Test icon (4.1) will indicate the status of the switch: