Seca 286 Bedienungsanleitung Und Garantieerklärung Seite 75

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
0,5 m
0,5 m
0,5 m
Foot screw
Air bubble
Knurled wheel
3. Affix the "Correct posture" label to the third column element.
NOTICE!
Malfunction caused by other ultrasonic emitters
If there are other ultrasonic emitters in the immediate vicinity of the
device - automatic door openers, for example - faulty measurements
or wireless transmission interference will occur.
Ensure that there are no other ultrasonic emitters in the same room
or in the immediate vicinity of the device.
WARNING!
Incorrect measurement due to reflections
If there are objects or people in the immediate vicinity of the device,
incorrect measurements will result.
Ensure that there are no objects or people within 0.5 meter of the
front or side of the scale during the measuring process.
Ensure that the device is at least 0.2 meter away from the wall.
Ensure that the patient is not wearing any kind of hair accessory on
top of the head.
4. Mark the area shown in the illustration using colored adhesive tape, for
example.
0,5 m
0,5
5. Align the device by turning the foot screws.
The air bubble in the spirit level must be located in the exact center of the
circle.
Spirit
level
6. Tighten the knurled wheels in the direction of the arrow.
The foot screws are secured against being adjusted.
Before you really get started ... • 75