5 Function and Assembly
5.1 Functional description
E
This device is made for temperature controlling of heatings, heated sample gas lines or heated systems.
Up to 4 heating circuits can be controlled.
The XT-T1 can be equipped with up to 4 controllers , each with a separate solid-state relay. Thus enables
a wear-free switching of the load current.
Therefore any numbers of cycle duties without wear and tear of contacts are possible.
Any control parameters will be set directly at the module. All controller parameters are set directly on the
module. The controller can be programmed both as a two- or three-point controller.
All common measuring transducers (e. g. PT100, Fe-CuNi, Ni-CrNi [standard]) can be connected.
Besides the connection plugs for the heating circuits the module is equipped with an additional alert
outlet.
When the display shows
I999
a sensor is broken. However, it may also be that the type of the sensor
has been entered incorrectly.
The actual value of the line temperature can be displayed on the screen and read even from a distance
due to the size and brightness of the display e.g. even outside of a test stand.
The control parameters are preset by us. But they may be altered depending on local needs. Read the
operating instructions carefully.
Product:
Temperature controller XT-T1
Revision 2.0 – Issue September 2018
96/166