Diagnosis and maintenance
460
19.2.7 Buffer overflow in command interface
Measured values lost due to buffer overflow in
catman®/Command interface.
19.2.8 System not ready
Device is damaged and not delivering valid measured
values.
Set to temporary during parameter set switching (ok).
Set to static if parameter set does not agree with card
configuration. Occurs when cards are removed/updated/
replaced or if an unsuitable parameter set was imported.
19.2.9 CPU overload during calculations
Computing time overrun in calculated channels. The
consequences can be gaps in the measured data flow.
Temporarily not harmful during parameter set switching,
editing of calculated channels or self-calibration of bridge
inputs.
If set during normal operation: Reduce number of
function blocks and/or reduce global sampling rate
(38.4kHz -> 19.2 kHz).
19.3 Resetting the PMX administrator
password
This procedure has no effect on the password for
maintenance work.