HBM PMX Bedienungsanleitung Seite 394

Vorschau ausblenden Andere Handbücher für PMX:
Inhaltsverzeichnis
PMX command set (API)
Syntax:
Parameters:
Note:
CAL?
Syntax:
Parameters:
Response:
392
CAL(x)
none
In all CF bridge amplifiers, this command triggers a
calibration. The measured values "flicker" for several
seconds.
The response depends on the SRB command.
Status of Calibration procedure
Output the status of the calibration procedure
The status of the calibration procedure, all selected
channels (PCS/SPS). Only supported for PX455! Other
(meas‐)cards return "OK"(0).
CAL?(x)
none
q1(y)
0
Automatic calibration is not executed
1
e.g. two boards with 4 channels each: 0,0,0,0:1,0,1,1
A3257-5.1 HBM: public
Automatic calibration is executed
PMX