PMX command set (API)
Response:
Syntax 2:
Parameters:
Response:
Syntax 3:
Response:
CAP
Syntax:
Parameters:
Effect:
390
q1,q2(y)
q1 Filter characteristics
q2 Limit frequency
e.g. 142,969:142,969
SFC??(x)
none
q1,q2(y): Possible filter characteristics
e.g. 141,142:141,142
SFC?142,?(x)
q1,...,qn(y): Available Bessel frequencies
e.g.
914,917,921,927,931,935,941,945,949,955,958,962,969,973
,1150:914,917...
Virtual subchannels (channel 9) do not support filters.
Query returns 140, 1150
Calibration Point
Enter the characteristic points of the transducer (input)
Affects all the selected channels (PCS/SPS)
CAPp1,p2,p3(x)
p1: Number of points (1 or 2)
p2: Measurement signal (amplifier-dependent unit), if
there is no input value present, the current measured
value is applied
p3: Display value
The input characteristic is defined by two points. The
input signal and the associated display value must be
entered for each point. This command defines also the
scaling of the PX878. That means the physical (p2, from
A3257-5.1 HBM: public