• Do not use the paper roll holder as a push/pull device. Product damage may occur.
• Do not hang items that exceed the safe working load of 1.5 kg on the paper roll holder.
• Always use caution when you attach restraint straps. Patient or operator injury may occur. Physical restraints, even if
secured, may result in serious harm to patients and operators, including entanglement, entrapment, physical injury, or
death.
• Always attach restraint straps or devices only at the identified attachment points of the product. Failure to do so may
result in patient or operator injury. Do not attach restraint straps to the siderail.
Always refer to applicable state and federal restrictions and regulations and the appropriate facility protocols before you
use any restraint strap or device.
Always refer to applicable state and federal restrictions and regulations for safety before you use the X-ray option with
radiation generating devices. Radiation generating devices may produce residual, stray, or scattered radiation.
• Always use caution when you take X-rays with the Fowler backrest in the upright position or when you use a lateral
cassette.
• Do not wash the internal components of this mattress. Discard the mattress if contamination is found inside.
• Do not immerse the mattress in cleaning or disinfectant solutions.
• Do not allow liquid to pool on the mattress.
• Do not iron, dry clean, or tumble dry the mattress cover.
• Do not clean, service, or perform maintenance while the product is in use.
• Do not steam clean, hose off, or ultrasonically clean the product. Use of these methods of cleaning is not recommended
and may void the product's warranty.
• Always disinfect the mattress following your hospital protocols to avoid the risk of cross-contamination and infection.
• Do not use V V i i r r e e x x ® ® T T B B to disinfect this product.
• Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the mattress
cover.
• Always inspect the mattress each time you clean the mattress cover. Follow your hospital protocols and complete
preventive maintenance each time you clean the mattress cover. If compromised, remove the mattress from use and
replace the product to prevent cross-contamination.
C C A A U U T T I I O O N N
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Always transport the S S T T 1 1 and S S T T 1 1 - - X X Series stretcher on floors made of wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30% to avoid electrostatic discharge.
• Do not use the hydraulics on the base to raise the product with a patient lift under the product.
• Do not place objects that exceed 60 lb (27 kg) in the base hood.
• Do not sit, step, or stand on the base hood.
• Always use authorized accessories with the S S T T 1 1 and S S T T 1 1 - - X X Series stretcher.
• Always use the 6300-1-000 mattress cover on the foam core.
• Always make sure that you wipe the product with clean water. Dry each product after cleaning. Some cleaning agents
are corrosive in nature and may cause damage to the product. Failure to follow these cleaning instructions may void
your warranty.
• Do not use cleaning agents and disinfectants with aggressive chemicals as they will reduce the expected life of the
mattress cover.
• Do not allow liquid to seep into the zipper area or watershed cover barrier when you clean the mattress bottom. Fluids
allowed to come in contact with the zipper may leak into the mattress.
• Always dry the mattress cover before you store, add linens, or place a patient on the surface. A dry product helps to
prevent impaired product performance.
• Do not overexpose the mattress cover to high concentrations of disinfectant solutions as they may degrade the mattress
cover.
• Failure to follow the manufacturing cleaning instructions and Stryker operational instructions may affect useful life of the
mattress.
EN
4
KK-6300 Rev 04