3. Move the standard rail at a distance of 20 mm from the top edge of the device
mount. Align this with a spirit level.
4. Tighten the two M10x65 hexagon socket screws (1).
Tightening torque: 30 Nm
4.3.2.3
Establishing potential equalization for the ceiling supply unit
1. Tighten the two M4x6 threaded pins (1) by hand.
4.4

Uncoupling from the Ambia ceiling supply unit

The following describes how to undo the standard rail for the Ambia ceiling supply
unit. The exact procedure for disconnecting the Atlan ceiling-mounted versions
from the ceiling supply unit is described in the supplement "Atlan A300, A350
ceiling-mounted versions and wall-mounted version".
Check that the floor under the device is level and cleared. There must be no
objects under the device.
|
Kit
Atlan A3xx SW 2.0n ceiling-mounted versions
Connecting Atlan to the Ambia ceiling supply unit
|