Herunterladen Inhalt Inhalt Diese Seite drucken

Coupling The Ceiling-Mounted Versions - Dräger Atlan A3-Serie Bedienungsanleitung

Anesthesia workstation
Inhaltsverzeichnis

Werbung

|
Connecting Atlan to the Ambia ceiling supply unit
4.3.2

Coupling the ceiling-mounted versions

4.3.2.1
Ceiling-mounted version with service cart
1. Connect Atlan to the ceiling supply unit.
2. Unscrew the M10x65 hexagon socket screws (1) on both sides of the frame rail
clamps of the standard rail.
3. Move the standard rail at a distance of 16 mm from the top edge of the device
mount. Align this with a spirit level (2).
4. Tighten the two M10x65 hexagon socket screws (1).
Tightening torque: 30 Nm
4.3.2.2
Ceiling-mounted version with integrated castors
1. Connect Atlan to the ceiling supply unit.
2. Unscrew the M10x65 hexagon socket screws (1) on both sides of the frame rail
clamps of the standard rail.