Using the unit

When used on the stand
Minimum clearance between the lower edge of photo-
therapy lamp and the patient = 30 cm.
Phototherapy in incubator
Example: Incubator 8000 NC
1 Loosen the clamp and adjust the height of the stand
for the intensity of light required (see table, p. 24).
Re-tighten the clamp.
Push the base of the stand under the incubator from
the rear until the lamp housing is centred over the
incubator.
2 Lock the castors (where applicable). The trolley can
no longer be moved.
Phototherapy in Warming Cot with radiant heater
Example: Babytherm 8000 OC
3 Loosen the clamp and adjust the height of the stand
for the intensity of light required (see table, p. 24).
Re-tighten the clamp.
Push the base of stand under the heated bed from the
rear.
4 Turn the setscrew, and tilt the Phototherapy Unit
(max. 30° possible) until the light shines on the side of
the patient.
Radiant heat from the heater must reach the patient
unimpeded.
5 Lock the castors (where applicable), so that the trolley
can no longer be moved.
When used without the stand
The lamp may also be placed on the canopy of the
incubator itself or on the canopy of the heated bed.
Check the load bearing capacity of the canopy.
Check the stability of the phototherapy lamp.
Minimum clearance to the patient = 30 cm.
Secure the lamp to prevent it from falling.
5
6
Fuse
7
minimum
30 cm
1
2
3
minimum
30 cm
4
5
Preparation
Using the unit
17