Herunterladen Inhalt Inhalt Diese Seite drucken

Parameter Access - HBM WE2108 Serie Referenzhandbuch

Teil 2
Vorschau ausblenden Andere Handbücher für WE2108 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

98
6
Overview:
(Define Password)
Defining a password
The user can enter an arbitrary, max. 7−digit password with this command. All ASCII charac-
ters which the WE2108 processes (see Section 2) are permitted. The entry must be in quota-
tion marks ("...").
(Set Password)
Write enable for password−protected user parameters
The command SPW with the correctly entered password authorizes entry of certain com-
mands. There is no distinction between uppercase and lowercase letters in the password
entry. No password is required for queries.
HBM

Parameter access

DPW
Determining a new password
SPW
Parameter enable with password
TDD
Storing / resetting the parameters
RDP
Direct parameter access (for service / experienced users)
WRP
Direct parameter access (for service / experienced users)
DPW
Command:
Parameter:
Range:
Factory setting:
Password protection: no
Parameter backup:
Query:
SPW
Command:
Parameter:
Range:
Factory setting:
Answer:
Password protection: −−
Parameter backup:
Query:
DPW ("passwrd")
1
1...7 letters or numbers (ASCII characters)
"WE8"
automatic
not intended
SPW("passwrd");
1
1...7 letters or numbers (ASCII characters)
−−
0CRLF with correct password
?CRLF with incorrect password
not possible
not intended
Scale electronic unit (NSW)
WE2108

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

We2108sWe2108m

Inhaltsverzeichnis