Scale electronic unit (NSW)
4
All commands which act directly on a measured value belong to measuring, these are:
NOTE
MSV and COF concern only the output of the measured value through the interface.
All other commands act in the same way on the display value in the display and on
the output value.
(Measured Signal Value)
Output measured values (format depending upon parameter COF)
The output measured value is related to the user characteristic and can be a gross or net
value (see command TAS). It is equal to the display value of the WE2108, but can be sent in
different formats. The output format for the measured value must be set previously through
the command COF. The length of the answer (number of characters) depends upon the set
format, but not upon the measured value.
The last generated measured value is stored in the WE2108 and is output on every MSV;
command. In the case of fast query, this can also be the same value two times, although the
weight is just changing.
WE2108

Measuring

MSV
COF
ASF
ICR
CDL
TAR
TAV
TAS
LIV Limit value setting
MSV
Command:
Query:
Answer:
Parameter:
Range:
Factory setting:
Password protection: no
Parameter backup:
Measured value output
Output format
Filter setting
Measuring rate setting
Zeroing
Taring
Set tare memory
Gross / net switch−over
not intended
MSV?;
outputs the current measured value
"1.35 V scale nominal value (ASCII / binary)
−−
not possible