88
(Loadcell Dead Load Weight)
Zero point of the user characteristic / scale characteristic
The command results in measurement and temporary storage of the internal value with the
scale unloaded. The value is offset only after the LWT command.
(Loadcell Weight)
End value of the user characteristic / scale characteristic
The command works analogously to the command LDW. The WE2108 calculates a new
characteristic from LWT and the previously determined value of LDW.
When calibrating with LDW / LWT, the parameter LVA2 must also be tested or set, since it is
not reset by LWT (see example).
HBM
LDW
Command:
Password protection: yes
Parameter backup:
Query:
Answer:
Parameter:
Range:
Factory setting:
LWT
Command:
Password protection: yes
Parameter backup:
Query:
Answer:
Parameter:
Range:
Factory setting:
LDW;
after entry of LWT
LDW?;
stored value, 7−digit with sign
1
0...270.000
0
LWT;
automatic
LWT?;
stored value, 7−digit with sign
1
0...270.000
200000