A low cutoff frequency means a good filter effect, but also a longer transient period when the
weight changes. The smallest filter setting at which the standstill is still guaranteed at con-
stant weight is optimum.
The stated transient periods refer to measured value acquisition by the WE2108. The total
transient period depends in addition upon the mechanical design and the dead load of the
scale and the weight to be weighed.
The output rate is not changed by ICR. High values of ICR result in a slower change of the
weight display after a load step change.
(Tare)
Performance of taring
The TAR command acts like pressing the tare key. The current measured value is tared, if
the standstill criterion is fulfilled. After successful taring, the unit switches over to "net mea-
sured value", this can be checked with TAS?;. The current value is filed in the tare memory
and is used for calculating all following net values.
(Tare Set)
Gross/net switch−over
Meaning of the parameter:
0 = Net. Scale display goes into net mode, on query with MSV?; the net value is output
1 = Gross. Scale display goes into gross mode, on query with MSV?; the gross value is out-
HBM
TAR
Command:
Parameter:
Password protection: no
Parameter backup:
Query:
TAS
Command:
Parameter:
Range:
Factory setting:
Password protection: no
Parameter backup:
Query:
put
TAR;
−−
−−
not intended
TAS(x);
1
0...1
1 (gross)
−−
TAS?;
Scale electronic unit (NSW)