HBM WE2108 Serie Referenzhandbuch Seite 84

Teil 2
Vorschau ausblenden Andere Handbücher für WE2108 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

84
(Clear dead load)
Zeroing the scale
The command CDL acts like pressing the zero key on the unit. Zeroing is performed if the
gross value lies in the allowed zeroing range and the standstill criterion is fulfilled.
(Limit Value) available from software P82 on
Limit value settings
Meaning and setting range of the parameters:
The meaning of these settings is described in detail in the WE2108 operating instructions.
HBM
CDL
Command:
Parameter:
Password protection: no
Parameter backup:
Query:
LIV
Command:
Parameter:
Range:
Factory setting:
Password protection: no
Parameter backup:
Query:
Command
Meaning
LIV 0,xx
Limit1 function
LIV 1,xx
Limit output 1 logic
LIV 2,xx
Switch−on value 1
LIV 3,xx
Switch−off value 1
LIV 4,xx
Limit2 function
LIV 5,xx
Limit output 2 logic
LIV 6,xx
Switch−on value 2
LIV 7,xx
Switch−off value 2
CDL;
−−
−−
not intended
LIV(0...7),(xx);
2
see tables
0
with command TDD1
LIV?(0...7);
Range
0: off
1: limit related to gross
2: limit related to net
0: active = open,
1: active = ground
0...CAP2
0...CAP2
s. LIV 0
0: active = open,
1: active = ground
0...CAP2
0...CAP2
Scale electronic unit (NSW)
equivalent menu item
P1_Fu
P1_Lo
P1_On
P1_OF
P2_Fu
P2_Lo
P2_On
P2_OF
WE2108