5.3
Calibration with calibration weights (zero and end value / partial load):
The procedure with and without partial load calibration does not differ. Partial load calibration
means that the calibration weight deviates from the scale nominal load.
NOTE
The commands LDW and LWT are performed only at standstill. Otherwise the answer
?CRLF is output and the value is not stored. If no standstill is reached, a very sensi-
tive characteristic has been set possibly by an earlier calibration. The WE2108 should
then be set to a standard characteristic with the following commands:
Entry of the values, with partial load calibration:
HBM

Examples

SPW"...";
Enter password
LVA2,5000
Value of the used calibration weight
ASF(...);
Set filter so that there is a stable display
With unloaded scale wait for standstill
LDW;
Acquire (and temporarily store) measured value
Load scale with calibration weight, wait for standstill
LWT;
Acquire and store measured value
ASF(...);
Set corresponding to the application
TDD1;
Store ASF nonvolatilely
SPW"...";
Enter password
LVA0,0;
Activate linear characteristic
LVA1, 0;
Internal value without load
LVA2,6000;
Characteristic point 1 display
LVA3,200000; Internal characteristic point 1 value
TDD1;
Nonvolatile storage of the characteristic
Subsequent calibration with LDW / LWT
Example: Scale nominal load 60 kg, 6000 divisions, i.e. display scaling 60.00.
Used calibration weight: 20 kg
³
SPW"...";
LVA0,0
LVA2,2000
LVA1,080000
LVA3,110000
TDD1;
Enter password
No linearization
Entry of the calibration weight
Measured value unloaded 0.8 mV/V
Measured value loaded (20 kg) 1.1 mV/V
Nonvolatile storage of the characteristic
Scale electronic unit (NSW)
WE2108