Error Rectification - HBM WE2108 Serie Referenzhandbuch

Teil 2
Vorschau ausblenden Andere Handbücher für WE2108 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Scale electronic unit (NSW)
5.5.3
If the geographical data were not entered correctly during the calibration, the scale indicates
incorrectly after the new entry of LGC1 / LGC2.
Possibly the setting with which the calibration was made is not known. If the calibration site is
known, the correct characteristic calculation can be ensured by the following entries:
Scale should be operated only at the calibration site:
After mV/V calibration with HBM load cells, installation site deviating from values for
Darmstadt:
Scale was calibrated in Darmstadt or the same use zone (see Section 5.5), installation
site deviating from values for Darmstadt:
Calibration site and installation site deviating from values for Darmstadt:
Switching off the correction
By the entry
the correction is temporarily deactivated (read old value previously with LGC?2; and note it).
Operation at the calibration site is basically possible with this setting. The correction is active
again after restoration of LGC2 ( <= 90).
WE2108

Error rectification

LGC0;
TDD1;
LGC0;
LGC1,xx;
xx = Altitude of the installation site/100
LGC2,xx;
xx = Latitude of the installation site
TDD1;
Proceed as according to mV/V calibration (see above)
LVA?0;
Read out and note all used characteristic values (see Section , LVA)
LVA?1;
LVA?.....
LGC1,xx;
xx = Altitude of the calibration site/100
LGC2,xx;
xx = Latitude of the calibration site
LDW;
Scale unloaded
LWT;
Scale loaded (>20% nominal value, accuracy immaterial)
LVA0,yyyy;
Enter noted characteristic values again (see Section , LVA)
LVA.....
LGC1,xx;
xx = Altitude of the installation site/100
LGC2,xx;
xx = Latitude of the installation site
TDD1;
LGC2,91;
TDD1;
97