Scale electronic unit (NSW)
5.5.3
If the geographical data were not entered correctly during the calibration, the scale indicates
incorrectly after the new entry of LGC1 / LGC2.
Possibly the setting with which the calibration was made is not known. If the calibration site is
known, the correct characteristic calculation can be ensured by the following entries:
Scale should be operated only at the calibration site:
After mV/V calibration with HBM load cells, installation site deviating from values for
Darmstadt:
Scale was calibrated in Darmstadt or the same use zone (see Section 5.5), installation
site deviating from values for Darmstadt:
Calibration site and installation site deviating from values for Darmstadt:
Switching off the correction
By the entry
the correction is temporarily deactivated (read old value previously with LGC?2; and note it).
Operation at the calibration site is basically possible with this setting. The correction is active
again after restoration of LGC2 ( <= 90).
WE2108
Error rectification
LGC0;
TDD1;
LGC0;
LGC1,xx;
xx = Altitude of the installation site/100
LGC2,xx;
xx = Latitude of the installation site
TDD1;
Proceed as according to mV/V calibration (see above)
LVA?0;
Read out and note all used characteristic values (see Section , LVA)
LVA?1;
LVA?.....
LGC1,xx;
xx = Altitude of the calibration site/100
LGC2,xx;
xx = Latitude of the calibration site
LDW;
Scale unloaded
LWT;
Scale loaded (>20% nominal value, accuracy immaterial)
LVA0,yyyy;
Enter noted characteristic values again (see Section , LVA)
LVA.....
LGC1,xx;
xx = Altitude of the installation site/100
LGC2,xx;
xx = Latitude of the installation site
TDD1;
LGC2,91;
TDD1;