Scale electronic unit (NSW)
Printout
5.5
The gravitation and thus the display of the scale depends upon the geographical data (lati-
tude and altitude above sea level) of the installation site. The change between different
regions is maximum 0.1 % within Germany and world wide 0.6 %.
A correction with the command LGC is recommended only if this error influence at the new
installation site exceeds the accuracy limits of the scale.
The commands should not be used in the case of on−site calibration
(see also: Error rectification).
The settings for Darmstadt may be used on the following cases:
WE2108
corresponds to
Adjustment
LIN
:
0
LOAd0 :
0
LOAd1 :
6000
LOAd2 :
0
LOAd3 :
0
CAL−0 :
0,30000
CAL−1 :
1,56000
CAL−2 :
0
CAL−3 :
0

Gravitation correction

For installation in the regions of Northrhine−Westfalia, Hesse, Rhineland−Palatinate,
Saarland
As "calibration site" for mV/V calibration with HBM load cells
LVA0
−− (no load, fixed)
LVA2
LVA4
LVA6
LVA1
LVA3
LVA5
LVA7