100
w
Parameters influenced by TDD..
HBM
The extent of the data stored with TDD1 depends upon the access authorization which
exists on execution of the command. Only the free parameters are stored without a pass-
word (limits, filters, interface parameters).
With set password TDD1 also stores the scale settings. With legal−for−trade mode
(setting Trade > 0) this causes the calibration counter to be incremented.
Certain commands (e.g. LWT) store the associated parameters automatically. TDD1 is
then not required (this is noted for the individual command descriptions). The following
applies for password−protected parameters which are not stored automatically (e.g.
LVA..): TDD1 stores the change only if the password is still set!
Factory setting
Setting after TDD0
ADR31
not changed
BDR7
not changed
COF9
not changed
E_Ch1 = 0Dhex
E_Ch1 = 0Dhex
E_Ch2 = 0Ahex
E_Ch2 = 0Ahex
St_Ch = 0
St_Ch = 0
TAS1
TAS1
TAV0
TAV0
Count = 5
Count = 5
P1... = 0
P1... = 0
P2... = 0
P2... = 0
IDNxxx..x,Nr,P8x
IDNxxx..x,Nr,P8x
PST"
"
PST"
F_Prt = 1
F_Prt = 1
SPACE = 0
SPACE = 0
LnEF1 = 0
LnEF1 = 0
LnEF1 = 2
LnEF1 = 2
ASF3
ASF3
ICR2
ICR2
DPW"WE8"
DPW"WE8"
LocPA = 0
LocPA = 0
unit = 2
unit = 2
Still = 1
Still = 1
Point = 2
Point = 2
res 1 = 1
res 1 = 1
LGC0
LGC0
CAP 1 = 6000
CAP 1 = 6000
LVA2,6000
LVA2,6000
LVA0,0
LVA0,0
LVA1,0
LVA1,0
LVA3,200000
LVA3,200000
Remarks
Address 31
9600 baud, even parity
ASCII measured value output, address, status
End character 1
End character 2
No start character
Gross measured value
Tare memory deleted
Reference quantity
Limits inactive, switching points = 0
Unit type, serial No., program version
"
Printer text line
Printing protocol
Printing format
Printing format
Printing format
Filter 8 Hz
Mean value over 2 measurements
Password
Parameter access free
Unit kg
1d / second
Places after the decimal point
Scale interval
Standard gravitation
Nominal value
Calibration weight
Linearization switched off
User characteristic zero point
User characteristic end value
Scale electronic unit (NSW)
WE2108