Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, and electromagnetic site survey should be
considered. If the measured field strength in the location in which the FlowOx™ is used exceeds the applicable RF
compliance level above, the FlowOx™ should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the FlowOx™.
Recommended separation distances between portable and mobile RF communications equipment and the FlowOx™
The FlowOx™ is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the FlowOx™ can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the FlowOx™ as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output
power
of transmitter W
0,01
0,1
0,2
1
1,8
2
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where E is the immunity test level
(V/m) and P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
610-00013 FLOWOX™ 2,0 Instructions for Use (IFU) Manual Rev 16
Separation distance according to frequency of transmitter m
380 MHz to 390 MHz
d=6/27* √ ��
0,03
0,08
0,10
0,23
0,30
-
6
�� =
∗ √��
��
430 MHz to 470 MHz
800 MHz to 2,57 GHz
d=6/28* √ ��
0,03
0,07
0,10
0,22
0,29
0,30
88
704 MHz to 787 MHz
5,1 GHz to 5,8 GHz
d=6/9* √ ��
0,07
0,21
0,30
-
-
-
Date of Issue: 28 APR 2022