Herunterladen Inhalt Inhalt Diese Seite drucken

Wenglor weCat3D MLSL-Serie Betriebsanleitung Seite 137

2d-/3d-profilsensoren
Inhaltsverzeichnis

Werbung

Byte Offset Tag Daten
Type:
21
0 = unsigned int
1 = float
22
Size in bits
23
Reserved
Element 2 of 4
ID-Name[0]:
0 = Dummy
1 = X
24
2 = Z
3 = Y
4 = I (Bit7-0: Int-Bit 10-2)
5 = Peak width (PW)
Type:
25
0 = unsigned int
26
Size in bits
27
Reserved
Element 3 of 4
ID-Name[0]:
0 = Dummy
1 = X
2 = Z
28
3 = Y
4 = I
5 = Peak width (PW)
5 = I-Low + PW(Bit7-6: Int-Low-
Bit1-0, Bit5-0: PW-Bit-5..0)
Type:
29
0 = insigned int
30
Size in bits
31
Reserved
Element 4 of 4
ID-Name[0]:
0 = Dummy
1 = X
2 = Z
32
3 = Y
4 = I
5 = Peak width (PW)
5 = I-height (Int-height-Bit7-2)
Type:
33
0 = unsigned int
34
Size in bits
35
Reserved
36
Reserved
2D-/3D-Profilsensoren
Beschreibung
0
16
Reserviert
4 = I
0
10
5 = PW
0
6
1 = X
0
16
Reserviert
Größe
Typ
in Byte
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
1
unsigned char
4
unsigned char[4]
137